Venosafe - Serum-Gel VF-052SAS

 
Пробирки Venosafe. Serum-Gel Product type: -
Serum-Gel
Catalogue number: -
VF-052SAS
Nominal volume: (ml)
5
Draw volume: (ml)
3
Additive: -
Gel+Clot activator
Concentration of additive: (mg/ml)
-
Dimensions - Length (mm)
75
Dimensions - Diameter (mm)
13
Stopper colour: -
Cherry red
Cap colour: -
Cherry red transparent
Shelflife: (month)
18
Centrifugation rate: (G/min)
1200/10 (+/-100)
Unit Box: pcs
100
Shipping carton: pcs
12x100
Measures Unit Box: LxWxH
363x222x304
 

Instructions for use:

• After blood collection, gently invert the tube several times to maximise the contact between blood and silica.
• Allow the blood to coagulate during 30 min.
• In emergency, one can reduce the clotting time, but there can be no guarantee for complete clotting. In case of uncompleted coagulation and due to the mechanical stimulation of the blood coagulation, fibrin formation may occur during centrifugation. This will be observed as a gelatinous mass in the serum.
• Centrifuge the tube at room temperature, during at least 10 min. with a relative centrifugal force of 1200 G (+/-100G). For optimal barrier formation a swing-out rotor is recommended. Although the tubes withstand 4000G (with stopper) the use of higher centrifugation force is not recommended. Higher centrifugation force might reduce the centrifugation time but will increase the amount of cells on top of the gel. Lower centrifugation force might result in an incomplete gel barrier.
• After complete separation with the gel barrier in place storage requirements depend on the stability of the analytes and are equivalent to storage requirements for decanted serum samples.

 


 